warning
🚧 This document is still being actively worked on and is subject to change. 🚧
| Authors |
|
| Responsible |
|
| Last Updated | 10/12/2025, 12:48:54 PM |
| Last Author | Kai Berszin |
Thermal Testing
Sensor Placement​
| Component | Node Nr | FDR Report Page Nr | Testing | Inflight |
|---|---|---|---|---|
| S-Band Antenna | 105 | 105 | Yes | Yes, one internal IC |
| UHf | 112 | 68 | Yes | No |
| COM Amateur Radio | 225 | 28 | Yes | Yes, 2 internal IC |
| COMs Tranceiver | 255 | 28 | Yes | Yes, 2 internal IC |
| OBC | 205 | 30 | Yes | Yes, 2 internal IC |
| Sensor Board | 315 | 68 | Yes | Yes, 2 internal IC |
| ADCS CubeMag | - | - | Yes | No, 2 internal IC |
| EPS Bottom Board | 705 | 27 | Yes | Yes 2 internal IC |
| EPS Top Board | 771,772 | 31 | Yes | Yes, 3 internal IC |
| EPS Battery Board | Yes | Yes 2 internal IC | ||
| Solar Panel - a | 1201 | 28 | Yes | Yes, 2 internal IC |
| Solar Panel - b | 1211 | 28 | Yes | Yes, 2 internal IC |
| Solar Panel - c | 1221 | 28 | Yes | Yes, 2 internal IC |
| Panel d | 1231 | 28 | Yes | No |
| Panel z+ | 1245 | 68 | Yes | No |
| Microfluidic Chip | 990 | 68 | Yes | Yes, 1 Thermistor |
| bioBox bottom | 1500,1504,1520,1604 | 28 | Yes | Yes, 4 Thermistors |
| bioBox up | 1524,1600,1620,1624 | 28 | Yes | Yes, 4 Thermistors |
| ADCS Cube | 512,655 | 32,28 | Yes | Yes, 2 interna IC |
| External Sensors | Yes, 9 or 11 for external measurements | No |
Sensor fixation​
This is a list of thermally conductive, adhesive epoxies that can be used to fixate the thermal sensors on the necessary components. All epoxies below pass NASA low outgassing tests.
EP21TCHT-1
- https://www.masterbond.com/tds/ep21tcht-1
- Service temperature range: 4 K to 400 F (ca. 204°C)
- Heat Cure schedule: Overnight at room temperature, followed by 2 hours at 175-200°F
- Website has a form to discuss application with the manufacturer and a techincal data sheet may be requested
EP53TC
- https://www.masterbond.com/newsrelease/ep53tc
- Service temperature range: -73°C to +149°C
- Heat Cure schedule: 2 hours at 80°C followed by 90-120 minutes at 125°C
- Website has a form to discuss application with the manufacturer and a techincal data sheet may be requested
EP29LPSPAO-1 Black
- https://www.masterbond.com/tds/ep29lpspao-1-black
- Service temperature range: 4 K to 250 F (ca. 121°C)
- Heat Cure schedule: 12-24 hours at room temperature, followed by heating
- Website has a form to discuss application with the manufacturer and a techincal data sheet may be requested
The decision was made to use tape, to not affect other tests and be able to use the components after the test.
- The side panel are planned to be coated in A276 or Second Surface Mirror (Silver/Inconel FEP SMM Radiator Tape from Dunmore Aerospace)
Procedure for Thermal Testing of Payload-Box​
Freezing Preparation​
- Put the ice into the Styrofoam cup so that the bottom of the cup is not visible.
- Pour cold tap water until the ice is barely covered.
- Insert and fix a thermometer into the cup.
- Stir at least every minute until the thermometer reaches 0-1°C, then start the test.
- Connect the thermocouples to the device. Start up the software.
Boiling Preparation​
- Calculate the boiling point of water based on atmospheric pressure reported by a local weather service.
- Pour 1.5 - 2.5 L of tap water into a pot on the stove.
- Put the water to boil.
- Lower the heat until no dangerous splashing.
- Wait 5 minutes.
- If the temperature is still at b.p. ±2°C, start the test. If not, adjust the heating, wait 2 more minutes, and repeat until b.p. ±2°C is reached.
Test Execution (applicable to freezing and boiling conditions)​
- If test readiness passed, stir the water and insert one thermocouple at the height of the reference thermometer. Fix the thermocouple.
- Wait until the reading stabilizes, i.e., within 0.1°C over the last minute.
- Stir every 10 seconds.
- Record the value.
- Repeat with every thermocouple connected to individual ports.
- If more ports than thermocouples exist, plug the most accurate thermocouple into other ports.
- Record all anomalies. If an outlier is detected, test the same thermocouple in a different port and another thermocouple in the same port. What to do when anomalies are detected? Can the code be adjusted during the testing?